Introduction
This type of gauge adopts stable heat treated hot filament that has tiny heat capacity as the vacuum sensor. It also has the function of automatic temperature compensation and dust protection. The gauge has the advantages of wide range, fast response, anti-pollution, oxidation resistance, small drift, long service life, stable performance and so on. Therefore, it is particularly suitable for rough and low vacuum measurements.
Main Technical Parameters
Measurement range | 1×105~1×10-1Pa | |
Filament material | Tungsten filament (platinum filament or nickel chrome wire is optional) | |
Cold resistance of filament | 86.5±1.7Ω(23±3℃) | |
Filament temperature | 40~300℃ | |
The scattered property | Less than ±25%(3×103~1×10-1Pa) | |
Vacuum connection mode and baking temperature | Φ15.5±0.5 Glass | 80℃ |
Φ15.5±0.1 Metal | 80℃ | |
KF10 Quick release flange | 80℃ | |
KF16 Quick release flange | 80℃ | |
KF25 Quick release flange | 80℃ | |
KF40 Quick release flange | 80℃ | |
CF16 Ultra high flange | 450℃ | |
CF25 Ultra high flange | 450℃ | |
CF35 Ultra high flange | 450℃ |