VGC50x Vacuum Gauge Controller

VGC50x Vacuum Gauge Controller

Sustainable solution for process measurement, control and data log INTRODUCTION Compatible to the wide range of INFICON active gauges, the new VGC50x series of active gauge controllers are able to monitor and data log the entire pressure range from 10⁻¹⁰ to 1500 mbar (10⁻¹⁰ to 1125 Torr) and the...
Chat Now

Product Details

Sustainable solution for process measurement, control and data log



INTRODUCTION


Compatible to the wide range of INFICON active gauges, the new VGC50x series of active gauge controllers are able to monitor and data log the entire pressure range from 10⁻¹⁰ to 1500 mbar (10⁻¹⁰ to 1125 Torr) and the set point status.



BENEFITS


◆ Simple operation with dot matrix menu guided display for parameter, sensor or general settings

◆ Very bright and clear Display for long distance Instrument read- out

◆ Bar graph display with setpoints's or pressure vs. time display

◆ Data log and parameter log function with USB port (rear side) and USB stick (front side)

◆ Ethernet interface

◆ Firmware upgrade available online or with USB stick

◆ Two free definable setpoints per channel with adjustable hysteresis

◆ High Resolution - 16 bit A/D converter

◆ Automatic identification of the INFICON active gauges

◆ Programmable 0 to 10 V Chart Recorder Output with logarithmic/ linear characteristics for single gauge or gauge combination (only VGC502 and VGC503)



SPECIFICATION


blob.png



Type
VGC501VGC502VGC503
Measurement channels
123
Display
LCD, LED backlightLCD, LED backlightLCD, LED backlight
Rate1/s101010
Connectable gauges with display range
CDG (A/D)Torr1×10-3×F.S. …1×F.S.1×10-3×F.S. …1×F.S.1×10-3×F.S. …1×F.S.
PCGmbar5×10-4…15005×10-4…15005×10-4…1500
PSGmbar5×10-4…10005×10-4…10005×10-4…1000
MPGmbar1×10-9…10001×10-9…10001×10-9…1000
PEGmbar1×10-9…1×10-21×10-9…1×10-21×10-9…1×10-2
MAGmbar1×10-9…1×10-21×10-9…1×10-21×10-9…1×10-2
BCGmbar5×10-10…15005×10-10…15005×10-10…1500
BPGmbar5×10-10…10005×10-10…10005×10-10…1000
HPGmbar2×10-6…10002×10-6…10002×10-6…1000

Measurement unit 

(selectable)


mbar, Torr, Pa, hPa, 

Micron, V

mbar, Torr, Pa, hPa, 

Micron, V

mbar, Torr, Pa, hPa, 

Micron, V

Setpoints
Setpoint relays
246
Channel assignment
11 or 21,2 or 3
Adjustment range
sensor dependentsensor dependentsensor dependent
Hysteresis
adjustableadjustableadjustable
Relay contact

floating changeover 

contact

floating changeover 

contact

floating changeover 

contact

Contact ratingV(ac)30 / 130 / 130 / 1
Connector
D-sub, 15-pin, maleD-sub, 25-pin, femaleD-sub, 25-pin, female
Analog output
RangeV0 ... 10.3, sensor analog output signal0 ... 10.3, sensor analog output signal0 ... 10.3, sensor analog output signal
Analog output
123

Programmable analog 

output


-11
Connector
D-Sub, 15-pin, maleD-Sub, 9-pin, maleD-Sub, 9-pin, male
Interface (digital)
Connector

USB slave, master and 

Ethernet, USB Typ A 

(stick), USB Type B, FCC68/RJ45

USB slave, master and 

Ethernet, USB Typ A 

(stick), USB Type B, FCC68/RJ45

USB slave, master and 

Ethernet, USB Typ A 

(stick), USB Type B, FCC68/RJ45

Power
SupplyV (ac)100 ... 240100 ... 240100 ... 240
FrequencyHz50 ... 6050 ... 6050 ... 60
ConsumptionW≤45≤65≤90

Operation temperature 

(ambiance)

°C+5 ... +50+5 ... +50+5 ... +50










































































blob.png